You don't need to be Editor-In-Chief to add or edit content to WikiDoc. You can begin to add to or edit text on this WikiDoc page by clicking on the edit button at the top of this page. Next enter or edit the information that you would like to appear here. Once you are done editing, scroll down and click the Save page button at the bottom of the page.
Please Take Over This Page and Apply to be Editor-In-Chief for this topic: There can be one or more than one Editor-In-Chief. You may also apply to be an Associate Editor-In-Chief of one of the subtopics below. Please mail us [1] to indicate your interest in serving either as an Editor-In-Chief of the entire topic or as an Associate Editor-In-Chief for a subtopic. Please be sure to attach your CV and or biographical sketch.
Field desorption (FD)/field ionization (FI) refers to an ion source for mass spectrometry first reported by Beckey in 1969.[1] In field ionization, a high-potential electric field is applied to an emitter with a sharp surface, such as a razor blade, or more commonly, a filament from which tiny "whiskers" have formed. This results in a very high electric field which can result in ionization of gaseous molecules of the analyte. Mass spectra produced by FI have little or no fragmentation. They are dominated by molecular radical cations M+. and less often, protonated molecules
.
In FD, the analyte is applied as a thin film directly to the emitter, or small crystals of solid materials are placed onto the emitter. Slow heating of the emitter then begins, by passing a high current through the emitter, which is maintained at a high potential (e.g. 5 kilovolts). As heating of the emitter continues, low-vapor-pressure materials get desorbed and ionized by alkali metal cation attachment.
Many earlier applications of FD/FI to analysis of polar and nonvolatile analytes such as polymers and biological molecules have largely been supplanted by newer ionization techniques. However, FD/FI remains one of the only ionization techniques that can produce simple mass spectra with molecular information from hydrocarbons and other particular analytes. The most commonly encountered application of FD/FI at the present time is the analysis of complex mixtures of hydrocarbons such as that found in petroleum fractions.
The recently developed liquid injection FD ionization (LIFDI) [1] technique "presents a major breakthrough for FD-MS of reactive analytes" [1]: Transition metal complexes are neutral and due to their reactivity, do not undergo protonation or ion attachment. They benefit from both: the soft FD ionization and the safe and simple LIFDI transfer of air/moisture sensitive analyte solution. This transfer occurs from the Schlenk flask to the FD emitter in the ion source through a fused silica capillary without breaking the vacuum.
References
Bibliography
- Alexandrov, M. L.; Gall, L. N.; Krasnov, N. V.; Nikolaev, V. I.; Pavlenko, V. A.; Shkurov, V. A., Dokl. Akad. Nauk SSSR 1984, 277, 379-383. (in Russian)
- Prókai, László (1990). Field desorption mass spectrometry. New York: M. Dekker. ISBN 0-8247-8303-4.
Acknowledgement and Attribution Regarding Sources of Content
Some of the initial content on this page may be incorporated in part from copyleft sources in the public domain including wikis such as Wikipedia and AskDrWiki. Drug information for patients came from the The National Library of Medicine. Infectious disease information may have come from the Centers for Disease Control (CDC). Differential Diagnoses are drawn from clinicians as well as an amalgamation of 3 sources: 1.The Disease Database; 2. Kahan, Scott, Smith, Ellen G. In A Page: Signs and Symptoms. Malden, Massachusetts: Blackwell Publishing, 2004:3; 3. Sailer, Christian, Wasner, Susanne. Differential Diagnosis Pocket. Hermosa Beach, CA: Borm Bruckmeir Publishing LLC, 2002:7 .