Quadrupole mass analyzer

You don't need to be Editor-In-Chief to add or edit content to WikiDoc. You can begin to add to or edit text on this WikiDoc page by clicking on the edit button at the top of this page. Next enter or edit the information that you would like to appear here. Once you are done editing, scroll down and click the Save page button at the bottom of the page.

Jump to: navigation, search

Please Take Over This Page and Apply to be Editor-In-Chief for this topic: There can be one or more than one Editor-In-Chief. You may also apply to be an Associate Editor-In-Chief of one of the subtopics below. Please mail us [1] to indicate your interest in serving either as an Editor-In-Chief of the entire topic or as an Associate Editor-In-Chief for a subtopic. Please be sure to attach your CV and or biographical sketch.

The quadrupole mass analyzer is one type of mass analyzer used in mass spectrometry. In a quadrupole mass spectrometer the quadrupole mass analyzer is the component of the instrument responsible for filtering sample ions, based on their mass-to-charge ratio (m/z). A quadrupole mass analyzer is essentially a mass filter that is capable of transmitting only the ion of choice. A mass spectrum is obtained by scanning through the mass range of interest over time.

How it works

Image:Paul Patent 2939952 Fig5.gif
Image from "Apparatus For Separating Charged Particles Of Different Specific Charges" Patent number: 2939952 [1]

The quadrupole consists of four parallel metal rods. Each opposing rod pair is connected together electrically and a radio frequency voltage is applied between one pair of rods, and the other. A direct current voltage is then superimposed on the R.F. voltage. Ions travel down the quadrupole in between the rods. Only ions of a certain m/z will reach the detector for a given ratio of voltages: other ions have unstable trajectories and will collide with the rods. This allows selection of a particular ion, or scanning by varying the voltages.

Applications

These types of mass spectrometers excel at applications where particular ions of interest are being studied because they can stay tuned on a single ion for extended periods of time. One place where this is useful is in liquid chromatography-mass spectrometry or gas chromatography-mass spectrometry where they serve as exceptionally high specificity detectors. Quadrupole instruments are often reasonably priced and make good multi-purpose instruments.

Triple quadrupoles

A linear series of three quadrupoles can be used; known as a triple quadrupole mass spectrometer. The first (Q1) and third (Q3) quadrupoles act as mass filters, and the middle (q2) quadrupole is employed as a collision cell. This collision cell is an RF only quadrupole (non-mass filtering) using He or N gas (~10-3 Torr, ~30 eV) to induce collisional dissociation of selected parent ion(s) from Q1. Subsequent fragments are passed through to Q3 where they may be filtered or scanned fully.

This process allows for the study of fragments (daughter ions) which are crucial in structural elucidation. For example, the Q1 may be set to "filter" for a drug ion of a known mass, which is fragmented in q2. The third quadrupole (Q3) can then be set to scan the entire m/z range, giving information on the sizes of the fragments made. Thus, the structure of the original ion can be deduced.

The arrangement of three quadrupoles was first developed by Jim Morrizon of LaTrobe University, Australia for the purpose of studying the photodissociation of gas-phase ions. Yet, the first triple-quadrupole for mass analysis was developed at Michigan State University by Dr. Christie Enke and graduate student Richard Yost in the late 1970's.

See also

References

External links

Template:Mass spectrometryit:Analizzatore di massa a quadrupolo

WikiDoc Help Menu

Quick Start..

Editing basics

Advanced editing

Communicating your edits

Help Videos You Can Watch


Acknowledgement and Attribution Regarding Sources of Content

Some of the initial content on this page may be incorporated in part from copyleft sources in the public domain including wikis such as Wikipedia and AskDrWiki. Drug information for patients came from the The National Library of Medicine. Infectious disease information may have come from the Centers for Disease Control (CDC). Differential Diagnoses are drawn from clinicians as well as an amalgamation of 3 sources: 1.The Disease Database; 2. Kahan, Scott, Smith, Ellen G. In A Page: Signs and Symptoms. Malden, Massachusetts: Blackwell Publishing, 2004:3; 3. Sailer, Christian, Wasner, Susanne. Differential Diagnosis Pocket. Hermosa Beach, CA: Borm Bruckmeir Publishing LLC, 2002:7 .

Personal tools
In other languages