Ultrasound Research Interface
You don't need to be Editor-In-Chief to add or edit content to WikiDoc. You can begin to add to or edit text on this WikiDoc page by clicking on the edit button at the top of this page. Next enter or edit the information that you would like to appear here. Once you are done editing, scroll down and click the Save page button at the bottom of the page.
An Ultrasound Research Interface (URI) is a software tool loaded onto a diagnostic clinical ultrasound device which provides functionality beyond typical clinical modes of operation. Before an ultrasound image can be displayed to the user, it must undergo a series of transformations, typically referred to as the ultrasound processing chain. As data moves through the processing chain, it may be desirable to acquire it at certain intervals, so it can be processed offline for research purposes. A set of processing parameters sent to the ultrasound device control how the processing chain performs its operations. It may also be desirable to change these processing parameters to achieve different imaging results. A clinical ultrasound user only has access to the ultrasound data in its final processed form, referred to as a B-Mode image. The format of the images are typically in the DICOM file format. The clinical user also has limited access to the processing parameters that can be modified; for reasons of device usability, and so the device cannot be put into a state that renders itself inoperable.
The URI provides a means to step beyond the limits of the clinical usage, and allow researchers to acquire data from certain points in the processing chain, as well as change a larger set of processing parameters.
Typical B-Mode Receive Processing Chain
A typical digital ultrasound processing chain for B-Mode imaging may look as follows:
- Multiple analog signals are acquired data from the ultrasound transducer (the transmitter/receiver applied to the patient)
- Analog signals may pass through one or more analog notch filters and a variable-gain amplifier (VCA)
- Multiple analog-to-digital converters convert the analog radio frequency (RF) signal to a digital RF signal sampled at a predetermined rate (typical ranges are from 20MHz to 160MHz) and at a predetermined number of bits (typical ranges are from 10 bits to 16 bits)
- Beamforming is applied to individual RF signals by applying time delays and summations as a function of time and transformed into a single RF signal
- The RF signal is run through one or more digital FIR or IIR filters to extract the most interesting parts of the signal given the clinical operation
- The filtered RF signal runs through an envelope detector and is log compressed into a grayscale format
Multiple signals processed in this way are lined up together and interpolated and rasterized into readable image.
Data Access
A URI may provide data access at many different stages of the processing chain, these include:
- Pre-beamformed digital RF data from individual channels
- Beamformed RF data
- Envelope detected data
- Interpolated image data
Where many diagnostic ultrasound devices have Doppler imaging modes for measuring blood flow, the URI may also provide access to Doppler related signal data, which can include:
- Demodulated (I/Q) data
- FFT spectral data
- Autocorrelated velocity color Doppler data
URI Tools
A URI may include many different tools for enabling the researcher to make better use of the device and the data captured, some of these tools include:
- Custom MATLAB programs for reading and processing signal and image data
- Software Development Kits (SDKs) for communicating with the URI, signal processing and other specialized modes of operation available on the URI
Further Reading
URI Links
- URI Users Group Website
- The Ultrasonix 500RP: a commercial ultrasound research interface
- Axius direct ultrasound research interface
- Sonix RP ultrasound research interface
Acknowledgement and Attribution Regarding Sources of Content
Some of the initial content on this page may be incorporated in part from copyleft sources in the public domain including wikis such as Wikipedia and AskDrWiki. Drug information for patients came from the The National Library of Medicine. Infectious disease information may have come from the Centers for Disease Control (CDC). Differential Diagnoses are drawn from clinicians as well as an amalgamation of 3 sources: 1.The Disease Database; 2. Kahan, Scott, Smith, Ellen G. In A Page: Signs and Symptoms. Malden, Massachusetts: Blackwell Publishing, 2004:3; 3. Sailer, Christian, Wasner, Susanne. Differential Diagnosis Pocket. Hermosa Beach, CA: Borm Bruckmeir Publishing LLC, 2002:7 .